The ability to acquire, display and interrogate three-dimensional volumes of image data has been well-established through various scientific disciplines. The medical field, in particular, has exposed the public to tomographic methods through now common medical procedures such as computed axial tomography (CAT), magnetic resonance imaging (MRI) and positron emission tomography (PET). In an analogous fashion, the focused ion beam (FIB) and scanning electron microscopy (SEM) can combine to generate tomographic data. While less common, the FIB-SEM tomographic method has demonstrated the ability to complete 3-D volumetric reconstruction at a resolution of 10 nm or better in all three dimensions. Thus, the method holds tremendous potential for materials and life science investigations. With the advent of simultaneous high-resolution SE imaging during the FIB sectioning process, it is possible to acquire several hundred tomographic SEM image slices a few nanometers (nm) “thick” in the span of less than one hour. This ability to rapidly acquire high density and high-resolution tomographic SEM slices at the nanoscale contributes to practical implementation of the FIB-SEM tomographic technique.
Step by stepThe process begins by defining a volume that will be FIB sectioned. Volumes are typically 200 to 500 um3, which translates into typical dimensions of 10-um wide, 5-um deep and 10 um in height. Suitable FIB milling currents range from 10 to 100 pA. Once milling of the volume is started the entire process is recorded continuously in real-time via AVI capture of the SEM image.Frames are captured every 10 to 20 sec with scan conditions that result in 10,000 to 60,000 electrons per pixel per frame, yielding a very good signal-to-noise ratio. Two frames recorded from one such capture are shown in Fig. 1. The sample is a precipitate phase from the heat affected zone of a stainless steel weld. The dark contrast is the precipitate phase and there is a platinum protective layer on the top. The inherent resolution of the technique is based primarily upon the interaction volume of the electron beam with the sample. The resolution of the technique is therefore in the range of 5 to 20 nm depending upon the SEM voltage and the material. Best results are obtained at lower voltage, 0.75 kV to 1.0 kV to maximize depth resolution. Latest updates
|
|
|
Use of this website is subject to its terms of use. Privacy Policy |